Central AsianResearch Index

VOLTAMPEROMETRIK ANALIZ ASOSIDA OZIQ-OVQAT MAHSULOTLARI TARKIBIDAGI BA’ZI OG‘IR METALL IONLARINI ANIQLASH

The article examines the analytical performance of mercury-film electrodes based on carbon–graphite materials for the determination of cadmium, lead and copper ions in food products and environmental samples by stripping voltammetry. A graphite-to-paraffin ratio of 50:50–60:40 was selected as the appropriate composition range. The conditions for in situ mercury-film formation and the metrological characteristics of the analytical procedure were investigated. In most cases, the relative error did not exceed 10–11%, while the minimum detectable concentrations for Cd, Pb and Cu were 0.02, 0.06 and 0.10 µg/mL, respectively. The applicability of planar electrodes for small-volume measurements in a drop mode was also demonstrated.

Сведения

Журнал
Agro ilm
Опубликовано
2026 · Т. 130, № Maxsus son 3
Язык
Узбекский
ISSN
2091-5616
Издатель
O'zbekiston qishloq va suv xo'jaligi LLC