ASSESSMENT OF THE QUALITY OF IMAGE MATCHING IN PET / CT AND ANALYSIS OF HYBRID IMAGING PROBLEMS IN RADIOTHERAPY PLANNING
The reliability of obtaining qualitative and quantitative information in positron emission tomography depends on the level of depth of quality control of the diagnostic image. Given the active application of the procedure for planning radiation therapy using positron emission tomography combined with computed tomography, an urgent task is to assess the accuracy of combining the modalities of diagnostic images. In this work, we estimated the discrepancy between the volumes of the radiopharmaceutical accumulation foci (when combining PET and CT modalities of images) using a computer-based radiation therapy planning system. The influence of additional factors that may affect the correct combination of images is considered: time-of-flight technology, the number of subsets (the number of free projections used during reconstruction). The functional dependences of the discrepancy between the volumes of modalities (PET and CT images) on the volume of potential foci of accumulation (which are used as hot spheres) of the radiopharmaceutical preparation are obtained. The discrepancy between the diameters of the spheres measured on a CT image with the real diameters of the spheres is estimated.