Determination of radio-chemical purity and specific activity of radionuclide lutetium-177 chloride without carrier
In this work, methods for determining the radiochemical purity and specific activity of the radionuclide lutetium-177 in the form of chloride are considered. These indicators were determined by thin layer and paper chromatography. As a mobile phase in TLC, a solution of sodium citrate with a concentration of 0.1 mol/l was used, and in paper chromatography a 10% solution of ammonium acetate and methanol in a phase ratio of 1:1; and NH4OH:CH3OH:H2O in a phase ratio of 0.4:2:4.
Сведения
- Опубликовано
- 2020 · Т. 24, № 1, С. 55-61
- Язык
- Английский
- ISSN
- 1025-8817 · 2181-077X
- Издатель
- Arifov Institute of Ion-Plasma and Laser Technologies of Uzbekistan Academy of Sciences